Skip to main content

Table 4 Common problems and troubleshooting guide.

From: Protocol: analytical methods for visualizing the indolic precursor network leading to auxin biosynthesis

Problem

Possible reasons

Solutions

Liquid does not pass through TopTips before loading plant samples

The slit on TopTips is too narrow

Increase the centrifugal force to make liquid pass through, or switch to a new TopTip

Liquid does not pass through TopTips after loading plant samples

Plant debris blocks the TopTip slit

Always try to avoid transferring plant debris into TopTips

Increase the centrifugal force to make liquid pass through

Use a dissecting probe to remove visible plant debris

Drift of LC retention time

Injection volume is too large

Evaporate eluate or extract to a smaller volume with vacuum concentrator

Reduce the sample volume injected onto the LC

Change of LC mobile phase

Check that correct, freshly prepared mobile phase solvents are being used

Purge the LC lines after changing solvents

LC solvents are blocked or leaking

Replace the precolumn filter

Check column connections and reinstall if needed

Check LC lines for leaking solvent

Low yield of labeled compounds, but normal yield of endogenous compounds and/or internal standards

Slow turnover

Increase the labeling time period

Insufficient intake of labeled precursors

Increase the concentration of labeled precursors in the labeling solution

Make sure plants are bathed in sufficient volume of labeling solution

Broad/tailed LC peaks

Injection volume is too large

Evaporate eluate or extract to a lower volume with vacuum concentrator

Reduce the sample volume injected onto the LC

The precolumn filter is dirty

Replace the filter

The column is dirty

Disconnect the LC line from the MS source and wash the column with solvent according to manufacturer’s instructions

Analyte m/z values do not match calculated values

The MS needs to be calibrated

Perform mass calibration with calibration standard mixture according to instrument manual

Reduced MS sensitivity

The MS needs to be calibrated

See above

The tune file needs to be modified

Adjust source parameters for optimal ion intensity using standard compounds