Skip to main content

Table 1 The problems and solution methods during the protocol

From: A label-free, fast and high-specificity technique for plant cell wall imaging and composition analysis

Number

Problem

Possible reason

Solution

1

A lack of good contiguous microsections

An inappropriate microtome was used

A suitable microtome should be selected based on the characteristics of the sample

2

Intensity is too low

Laser intensity is too low

Make sure the intensity of the laser at the sample position is correct; increase laser power accordingly

 

Exposure time is insufficient

Increase the exposure time

 

Samples are overtreated and not suitable for imaging

Prepare the samples carefully according to step 1

3

Poor spectral resolution of the CRS spectrum

The sections of the sample are too thick, or the slit is too wide

Make sure the sections are ≤ 10 μm thick and that the slit opening is ≥ 50 μm

4

Too much fluorescence of background

Coverslips are dirty, there are bubbles, or the samples are damaged

Use new coverslips or prepare a new sample for imaging

5

Other fluorescence interference

Chloroplast auto-fluorescence

The sections of the sample can be immersed in 75% ethanol before spreading onto the slide

6

Sample burns

The sections of the sample were embedded in LR White resin or laser intensity is too high

Both herbaceous and woody samples should be prepared from fresh or dry plant material but not embedded

The laser intensity should be sufficiently high to obtain good counts per molecule (CPM), to ensure a good signal-to-noise ratio, but also sufficiently low to avoid photobleaching