Table 6 Comparison of the methods used to calculate microspore density. Accuracy is expressed in terms of deviation from the real/inicial value, precision in terms of coefficient of variation (CV) and reproducibility in terms of coefficient of repeatability (CR)

From: Comparison of six different methods to calculate cell densities

 

Accuracy (% deviation)

Precision (CV)

Reproducibility (CR)

Approximate time needed (min)

Ease of use

Estimated round basic price

Strength

Weakness

Neubauer

High (13%)

High (17%)

Moderate (175.9)

45

Very easy

€ 260 + microscope

Cheap, straightforward and accurate

Only good precision with high number of countings

Flow cytometry + PI

Low (32%)

Low (26%)

High (39.2)

90a

Moderate

Highly reproducible

Slow, expensive, low accuracy and precision

~€35.000 (optional ~ €29.000)

Flow cytometry + SSC

Low (30%)

Low (27%)

High (36.3)

15

Easy

~€35.000

Fast, highly reproducible, straightforward

Expensive, low accuracy and precision

Automated counter

Low (32%)

Moderate (19%)

Moderate (109.6)

15

Very easy

€15.000–€20.000

Fast and easy

Moderately expensive and accurate

Field counting

Low (33%)

Moderate (21%)

Low (248.4)

30

Easy

Inverted microscope

Very cheap and straightforward

Laborious, low accuracy and reproducibility

  1. Estimated round prices are given for a Partec Cyflow cytometer excluding taxes
  2. aIncluding the duration of incubation with PI